
Total Hydrocarbon (THC) analysis is essential to identify volatile and semi-volatile organic compounds (VOCs) that may leach from medical devices into the patient environment. These compounds may originate from:
Lubricants, silicone oils, mold release agents
Residual solvents (IPA, hexane, toluene, xylene)
Ethylene oxide (EO) sterilization by-products
Packaging material emissions
PVC, PU, PE, PP, silicone, TPU polymer degradation products
THC analysis is directly related to ISO 10993-18 (Chemical Characterization) and ISO 10993-17 (Toxicological Risk Assessment).
| Standard / Regulation | Scope |
|---|---|
| ISO 10993-18 | Chemical characterization of medical device materials |
| ISO 10993-12 | Sample extraction (surface area/volume ratio, solvent selection) |
| ISO 10993-5 | Cytotoxicity testing of extracts |
| USP <467> | Residual solvent limits |
| TS EN 12619 / EPA 25A | THC/VOC measurement using Flame Ionization Detector (FID) |
| ISO 13485 / FDA 21 CFR 820 | Process control and documentation |
FID detects organic compounds containing C–H bonds by combusting them in a hydrogen-air flame, producing ions. The ion current is proportional to the hydrocarbon concentration.
| Parameter | Description |
|---|---|
| Flame Gas | Hydrogen + Synthetic Air |
| Detection Limit (LOD) | ~0.05 ppmC |
| Quantitation Limit (LOQ) | ~0.1 ppmC |
| Output Units | ppmC (carbon equivalent), mg/m³ |
| Parameter | Specification |
|---|---|
| Extraction Ratio | 6 cm²/mL (surface area-based extraction) |
| Extraction Media | 0.9% NaCl, Ethanol 70%, PBS, IPA |
| Extraction Temperature | 37°C or accelerated at 50°C |
| Duration | 24–72 hours |
| Filtration | 0.22 μm PTFE filter (low VOC adsorption) |
| Setting | Value |
|---|---|
| Column | DB-624 (30 m x 0.32 mm x 1.8 μm) |
| Carrier Gas | Helium (1.5 mL/min) |
| Injection Volume | 1 μL (Split 1:20) |
| Injector Temp. | 200°C |
| Oven Program | 40°C → 200°C at 10°C/min |
| Detector (FID) Temp. | 250°C |
| Hydrogen Flow | 40 mL/min |
| Air Flow | 450 mL/min |
| Gas Standard | Concentration | Certification |
|---|---|---|
| Propane (C₃H₈) | 1% ±1% | ISO 6142 certified |
| Methane (CH₄) | 1% | Alternative carbon reference |
| Zero Air | Hydrocarbon-free | Baseline correction |
Example:
Measured Value: 20 ppmC (Propane equivalent)
| Parameter | Limit | Reference |
|---|---|---|
| Total Hydrocarbon Residue | < 10 mg/device | ISO 10993-12 & FDA |
| Residual Solvents (IPA, Hexane) | < 500 µg/device | USP <467> |
| THC in Sterilization Chamber | < 50 ppmC | TS EN 12619 |
| EO Sterilization VOC Residue | < 250 µg/device | ISO 11135 |
✔ System Suitability Test (SST) performed before analysis
✔ Blank measurement (zero air) < 0.1 ppmC
✔ Linearity: R² ≥ 0.995
✔ Repeatability: RSD < 5%
✔ Recovery: 80–120% (spiked QC samples)
✔ Positive Control: Heptane or Toluene standard solution
A compliant test report must include:
Sample description (material, batch number, sterilization type)
Extraction conditions (time, temperature, solvent, SA/V ratio)
GC-FID chromatograms and calibration curves
Raw THC values (ppmC) and converted values (mg/m³, µg/device)
Comparison with ISO/FDA limit values
Risk assessment according to ISO 10993-17
Conclusion: Complies / Does Not Comply