Pre-Treatment: Extraction • Determination: Gravimetric Method
ISO 2271 • ASTM F2459
During cleaning or reprocessing of medical devices, detergents containing anionic surfactants (e.g., SDS, SLES) may remain on the device surface. These residues may cause:
Cytotoxicity (cell membrane disruption)
Protein denaturation and inflammation
Interference with sterilization efficacy
Non-compliance with ISO 10993 biocompatibility requirements
| Standard | Description |
|---|---|
| ISO 2271 | Determination of anionic surfactants (gravimetric or MBAS method) |
| ASTM F2459 | Extractable detergent analysis from medical device surfaces |
| ISO 10993-12 | Sample preparation and extraction conditions |
| ISO 17664 / ISO 15883 | Cleaning and decontamination requirements |
| Parameter | Typical Condition |
|---|---|
| Solvent | Ultrapure water or 70:30 water:isopropanol |
| Temperature | 37°C–50°C |
| Duration | 30–60 minutes |
| Method | Stirring or ultrasonic extraction |
| Surface/Volume Ratio | 1 cm² = 1 mL solvent |
The extract is transferred into a pre-weighed dish
Solvent is evaporated at 105 °C
Residue is weighed
Result is expressed as mg/device or µg/cm²:
| Unit | Limit |
|---|---|
| mg/device | ≤ 0.5 mg |
| µg/cm² | ≤ 2 µg/cm² |
| Visual check | “No visible residue” (ASTM F2459) |
✔ ISO 2271 gravimetric & MBAS analysis
✔ ASTM F2459 compliant extraction & validation
✔ Post-sterilization surface cleanliness verification
✔ Toxicological risk evaluation (ISO 10993-17)
✔ Documented reports for CE / FDA compliance