According to TS 8195 / EN 1484**
TOC measures the total amount of organic carbon in a liquid. It is widely used to evaluate:
✔ Cleanliness validation of medical devices
✔ Residual organic contaminants after washing/sterilization
✔ Water quality used in production and rinsing (DI / WFI)
| Standard | Description |
|---|---|
| TS 8195 / EN 1484 | Guidelines for determination of Total Organic Carbon in water |
| ISO 10993-12 | Sample preparation and extraction conditions |
| ISO 15883 & ISO 17664 | Medical device cleaning validation requirements |
All organic compounds are oxidized to CO₂ via combustion or UV-persulfate oxidation.
The CO₂ is quantified by a NDIR detector.
TOC is calculated as:
(Where TC = Total Carbon, IC = Inorganic Carbon)
| Parameter | Condition |
|---|---|
| Extraction Medium | TOC-free ultrapure water |
| Surface/Volume | 1 cm² per 1 mL (ISO 10993-12) |
| Temperature | 25–37 °C |
| Duration | 30–72 h |
| Sample Type | Limit Value |
|---|---|
| Extract from medical devices | ≤ 2.0 mg/L |
| Final rinse water of cleaning system | ≤ 0.5 mg/L |
| Purified/WFI water | ≤ 0.1 mg/L |
✔ TOC analysis per TS 8195 / EN 1484
✔ Cleaning validation of surgical instruments and implants
✔ EO, gamma sterilization residue monitoring
✔ ISO 10993 & ISO 13485 compliant documentation
✔ Calibration, blank control, quality assurance